Sanitary Ware & Plumbing TestingCeramics
Sanitary Ceramic Crazing Resistance Tester
The instrument tests crazing resistance of sanitary ceramic ware using controlled hot- and cold-water baths. It references GB/T 6952-2015 and provides nickel-chromium heating, compressor cooling, circulation, automatic temperature control, and automatic stop at test completion.
- Model
- TKL-B
- Category
- Sanitary Ware & Plumbing Testing
Product description
Product Description
The hot bath is heated by nickel-chromium resistance wire and the cold bath is cooled by a compressor. A circulation pump draws from the upper cold-bath region and returns water at the lower region to circulate the bath.
Key parameters include bath temperatures, power, bath dimensions, FRPP tank construction, and copper pump construction.
Specifications
Technical Specifications
| Parameter | Specification |
|---|---|
| Referenced standard | GB/T 6952-2015 |
| Hot-bath temperature | 110 °C ±5 °C |
| Cold-bath temperature | 2.5 °C ±0.5 °C |
| Power supply | AC 220 V |
| Input power | 2.5 kW |
| Hot-bath dimensions | Ø250 mm × 325 mm |
| Cold-bath dimensions | Ø250 mm × 385 mm |
| Hot-bath heating | Nickel-chromium resistance wire |
| Cold-bath cooling | Compressor |
| Tanks | FRPP water tank; copper pump |
| Control | Digital regulator and time relay; automatic temperature control; automatic stop |