Glass TestingGlass
Glass Softening Point Tester by Hanging-Fiber Method
The RH-I determines glass softening temperature by the hanging-fiber method specified in ISO 7884-1978(E). It controls heating at 5 ± 1 °C/min and detects the temperature at which a 235 mm fiber elongates by 1 mm, corresponding to a viscosity of 4.5 × 10⁶ Pa·s.
- Model
- RH-I
- Category
- Glass Testing
Product description
Product Description
A glass fiber is suspended in the furnace and heated at a controlled rate. The instrument records the temperature at which the fiber reaches the specified elongation.
The system provides digital temperature, micrometer, and timing measurements. The instrument provides a maximum use temperature of 900 °C and an instantaneous maximum of 1,100 °C.
Specifications
Technical Specifications
| Parameter | Specification |
|---|---|
| Referenced standard | ISO 7884-1978(E); ISO 7884-1978 |
| Heating rate | 5 ± 1 °C/min |
| Elongation rate at softening | 1 mm |
| Corresponding viscosity | 4.5 × 10⁶ Pa·s |
| Glass-fiber length | 235 mm |
| Heating power | 600 W |
| Maximum operating temperature | 900 °C continuous; 1,100 °C instantaneous |
| Temperature resolution | 0.1 °C |
| Micrometer resolution | 0.01 mm |
| Timing accuracy | 0.1 s |
Inquiry
Contact Us for Pricing
Interested in this product? Reach out for a personalized quote or more information and configuration guidance.